Filament Maker TWO sensor calibration
The Filament sensor can become dusty or in rare instances misaligned, so to verify its accuracy, there is a dedicated sensor calibration procedure.
The sensor has to be calibrated in the following cases:
- The measured filament diameter does not equal to the physical diameter of the produced filament.
- The sensor's performance should be checked every year, as detailed in the Maintenance Schedule.
This procedure is carried out in house before shipping and the calibration kit is delivered with the Filament Maker TWO.